• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KINETIC CONCEPTS, INC. ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM; OMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KINETIC CONCEPTS, INC. ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM; OMP Back to Search Results
Model Number WNDARM
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 11/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on information provided, it cannot be determined that the alleged bleeding event requiring hospitalization and transfusion is related to the activ.A.C.¿ ion progress¿ remote therapy monitoring system.The patient was on anticoagulant therapy.The device passed quality control checks before and after patient placement.Device labeling, available in print and online, states: contraindications- do not place foam dressings of the v.A.C.® therapy system directly in contact with exposed blood vessels, anastomotic sites, organs, or nerves.Warnings- bleeding: with or without using v.A.C.® therapy, certain patients are at high risk of bleeding complications.The following types of patients are at increased risk of bleeding, which, if uncontrolled, could be potentially fatal: patients who have weakened or friable blood vessels or organs in or around the wound as a result of, but not limited to: suturing of the blood vessel (native anastomosis or grafts) / organ, infection, trauma, radiation, patients without adequate wound hemostasis, patients who have been administered anticoagulants or platelet aggregation inhibitors, patients who do not have adequate tissue coverage over vascular structures.If v.A.C.® therapy is prescribed for patients who have an increased risk of bleeding complications, they should be treated and monitored in a care setting deemed appropriate by the treating physician.If active bleeding develops suddenly or in large amounts during v.A.C.® therapy, or if frank (bright red) blood is seen in the tubing or in the canister, immediately stop v.A.C.® therapy, leave dressing in place, take measures to stop the bleeding and seek immediate medical assistance.The v.A.C.® therapy units and dressings should not be used to prevent, minimize or stop vascular bleeding.Protect vessels and organs: all exposed or superficial vessels and organs in or around the wound must be completely covered and protected prior to the administration of v.A.C.® therapy.Always ensure that v.A.C.® foam dressings do not come in contact with vessels or organs.Use a thick layer of natural tissue should provide the most effective protection.If a thick layer of natural tissue is not available or is not surgically possible, multiple layers of non-adherent dressing material may be considered as an alternative, if deemed by the treating physician to provide a complete protective barrier.If using non-adherent materials, ensure they are secured in a manner that will maintain their protective position throughout therapy.Dressing changes: wounds being treated with the v.A.C.® therapy system should be monitored on a regular basis.In a monitored, non-infected wound, v.A.C.® dressings should be changed every 48-72 hours, but no less than 3 times a week, with frequency adjusted by the clinician as appropriate.Infected wounds must be monitored often and very closely.For these wounds, dressings may need to be changed more often than 48-72 hours; the dressing changing intervals should be based on a continuing evaluation of the wound condition and the patient's clinical presentation, rather than a fixed schedule.Disclaimer: this information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a kinetic concepts, inc.Product malfunctioned, is defective or has caused serious injury.
 
Event or Problem Description
On 22-nov-2022, the following information was provided to kci by the patient: the activ.A.C.¿ ion progress¿ remote therapy monitoring system was removed allegedly due to bleeding from his wound.On (b)(6) 2022, the following information was provided to kci by the home health nurse: upon arriving at the patient's home, there was a pool of blood underneath the patient's foot.The nurse described the alleged bleeding event as shooting out of the wound.The nurse stated the patient was admitted to the hospital and received a blood transfusion.The nurse believed the bleeding event was due to the procedure done on (b)(6) 2022, but was not certain.On (b)(6) 2022, the following information was provided to kci by the triage nurse: the nurse confirmed the patient was hospitalized for bleeding and received a blood transfusion.No additional information was provided.On (b)(6) 2022, a device evaluation was completed by kci quality engineering.On (b)(6) 2022, the device was tested per quality control procedure by kci service center, and the unit passed the quality control checks and met specifications.On (b)(6) 2022, the device was placed with the patient.On (b)(6) 2022, the device was tested per quality control procedure by kci service center, and the unit passed the quality control checks and met specifications.Inspection and testing of the device did not reveal any evidence of an operational malfunction with the unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM
Common Device Name
OMP
Manufacturer (Section D)
KINETIC CONCEPTS, INC.
san antonio TX 78249
Manufacturer Contact
steven jackson
6203 farinon drive
san antonio, TX 78249
2102556438
MDR Report Key15996101
Report Number3009897021-2022-00177
Device Sequence Number5636233
Product Code OMP
UDI-Device Identifier00849554005600
UDI-Public0100849554005600
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K201571
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberWNDARM
Device Catalogue Number420095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2022
Initial Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALBUTEROL 90 MCG/ACTUATION INHALER, 2 PUFFS INHALE.; ALLOPURINOL 100 MG TABLETS, 1 TABLET BY MOUTH ONCE.; AMLODIPINE 5 MG TABLETS, 1 TABLET BY MOUTH ONCE DA.; ASPIRIN EC 81 MG TABLETS, 1 TABLET BY MOUTH ONCE D.; ATORVASTATIN 80 MG TABLETS, 40 MG TABLET BY MOUTH.; DULOXETINE 30 MG DELAYED RELEASE CAPSULE, 1 CAPSUL.; FLUTICASONE PROPIONATE 50 MCG/ACTUATION, 1 SPRAY I.; GABAPENTIN 300 MG CAPSULES, 1 CAPSULE BY MOUTH THR.; HYDROMORPHONE 2 MG TABLETS, 1 TABLET BY MOUTH EVER.; HYDROXYZINE 10 MG TABLETS, 1 TABLET BY MOUTH THREE.; INSULIN HUMAN U-100 NPH-REGULAR 70-30 MIX 100 UNIT.; ISOSORBIDE MONONITRATE CR 30 MG 24-HOUR TABLET, 1.; METOPROLOL XL 50 MG 24-HOUR TABLET, 1 TABLET BY MO.; MULTIVITAMIN TABLETS, 1 TABLET BY MOUTH ONCE DAILY.; NITROGLYCERIN 0.4 MG SL, DISSOLVE 0.4 MG UNDER TON.; OMEGA-3 FATTY ACIDS 340-1000 MG CAPSULES, 1 CAPSUL.; OMEPRAZOLE 20 MG DELAYED RELEASE CAPSULE, 1 CAPSUL.; TICAGRELOR 90MG TABLETS, 1 TABLET BY MOUTH ONCE DA.; TRAZODONE 100 MG TABLETS, 1 TABLET BY MOUTH ONCE D.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexMale
Patient Weight77 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-