CARDINAL HEALTH UMBILICAL CATHETER SINGLE LUMEN 5FR; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number 8888160341
Medical Device Problem Codes Break (1069); Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 12/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that the patient was being held when the family noticed blood backing up into the line and dripping outside of the tubing suddenly.She hit her call light and the staff came to the bedside to assess.There was a crack in the umbilical venous catheter (uac) lumen.This crack was not present earlier in the shift as the staff had been using the line for labs.The staff notified the advanced practice nurse (apn) who then came to bedside to remove the line.There was no patient harm reported.Additional information was received and stated that it being a very slow drip, might be a total of 1-2cc of blood dripped before she was able to get to bedside to remove the line.Per customer, the device was removed on (b)(6) 2022, and was not replaced.
 
Event or Problem Description
The customer reported that the patient was being held when the family noticed blood backing up into the line and dripping outside of the tubing suddenly.She hit her call light and the staff came to the bedside to assess.There was a crack in the umbilical venous catheter (uac) lumen.This crack was not present earlier in the shift as the staff had been using the line for labs.The staff notified the advanced practice nurse (apn) who then came to bedside to remove the line.There was no patient harm reported.Additional information was received and stated that it being a very slow drip, might be a total of 1-2cc of blood dripped before she was able to get to bedside to remove the line.Per customer, the device was removed on (b)(6) 2022, and was not replaced.Further information was received and stated that per customer, povidone iodine was used to clean the skin area prior to insertion.The line was secured with tegaderm and the ¿goal post¿ securement device.The device was inserted on (b)(6) 2022.The customer further stated that the lines would have been flushed with labs or if the patient received intermittent iv medications.They also run heparinized fluids through each lumen to keep lines patent.Per customer, 1 ml of normal saline in a 10 ml syringe was used to flush each time.The patient has been discharged from the hospital.
 
Additional Manufacturer Narrative
Section d6a / d6b field has been updated to include the device implant and explant date.Section b5 has been updated to include additional information.
 
Additional Manufacturer Narrative
Since a lot number was not provided, the device history record review could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.One used catheter from an unknown lot number was received inside a generic plastic bag for investigation.Upon visual evaluation, the catheter did not show signs of use.The physical sample evaluation was performed and revealed that the sample showed bubbles during underwater test.The catheter showed on its extension a crack where the air escapes and bubbles were identified.Therefore, the reported issue is confirmed.Based on the analysis, the issue was identified during the physical evaluation of the patient and the incident description states as ¿this crack was not present earlier in the shift as the staff had been using the line for labs¿.Based on the investigation, it was determined that the catheter was in good conditions prior to use.Also, manufacturing process was reviewed and no potential causes contributing to the reported issue were found.Based on the review of instructions for use (ifu) and replica done, it was determined that not correctly following the instruction for use is the most probable root cause of the reported issue.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UMBILICAL CATHETER SINGLE LUMEN 5FR
Common Device Name
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
edificio b20 calle #2 zona fra
alajuela
CS  
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15996144
Report Number3009211636-2022-00803
Device Sequence Number17632773
Product Code FOS
UDI-Device Identifier10884527005123
UDI-Public20884527005120
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8888160341
Device Catalogue Number8888160341
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2022
Supplement Date Received by Manufacturer12/08/2022
12/08/2022
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date12/19/2022
01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient Age1 WK
Patient SexFemale
-
-