| Model Number |
11061640 |
| Medical Device Problem Codes |
Computer Software Problem (1112); Loss of Data (2903)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens has completed an investigation of the event.Assessment of the log files does not indicate a system failure or malfunction and no non-conformity was identified.Investigation revealed that the system worked as specified.The customer had deleted the raw data before continuing the reconstruction.Based on the available information no general design issue was identified.
|
| |
|
Event or Problem Description
|
|
It was reported to siemens that a malfunction occurred while operating the somatom go.Top ct system.The customer reported that the images of a cta head stroke protocol study performed on (b)(6) 2022, could not be reconstructed on (b)(6) 2022.The patient was scanned with two spirals.Both raw data files did not later exist in the image reconstruction system (irs).The patient is a pregnant female (approx.Week 32 at time of scan) and was not rescanned.No negative health consequences were reported.We consider pregnant patients as a special population group who needs to be protected.Therefore, this report is filed with an abundance of caution.
|
| |
|