• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; TIBIAL POLY SPACER, 10MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; TIBIAL POLY SPACER, 10MM Back to Search Results
Model Number 25001210E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were revised: distal femur, two male-female midsections, tibial hinge component, poly spacer, and distal femur axial pin.The date of the patient's index eleos surgery is unknown and the implants placed during this procedure are unknown.Therefore, it is unknown if there were additional implants at the time of this patient's alleged infection.Attempts were made and no further information was made available.
 
Additional Manufacturer Narrative
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this event: #3013450937-2022-00383, #3013450937-2022-00384, #3013450937-2022-00385, #3013450937-2022-00386, #3013450937-2022-00388.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records could not be reviewed as the product information of the devices involved in the adverse event are unknown.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: b4: date of this report added g3: date received by manufacturer added g6: type of report added h2: follow-up type added h3: device evaluated by manufacturer updated to no h6: type of investigation code updated to 4111: communication/interviews h6: type of investigation code updated to 4110: trend analysis h6: type of investigation code updated to 4114: device not returned h6: type of investigation code updated to 4117: device not accessible for testing h6: investigation findings code updated to 3221: no findings available h6: investigation conclusions code updated to 4315: cause not established h10: additional narratives/data.
 
Event or Problem Description
It was reported that the patient developed an alleged infection.The patient underwent a revision surgery on (b)(6) 2022 where the following eleos implants were revised: distal femur, two male-female midsections, tibial hinge component, poly spacer, and distal femur axial pin.The date of the patient's index eleos surgery is unknown and the implants placed during this procedure are unknown.Therefore, it is unknown if there were additional implants at the time of this patient's alleged infection.Attempts were made and no further information was made available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
TIBIAL POLY SPACER, 10MM
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
MICROPORT ORTHOPEDICS
5677 airline road
arlington TN 38002
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key15996371
Report Number3013450937-2022-00387
Device Sequence Number5922524
Product Code KRO
UDI-Device IdentifierB27825001210E0
UDI-PublicB27825001210E0
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25001210E
Device Catalogue Number25001210E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2022
Supplement Date Received by Manufacturer02/16/2023
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000007E, ELEOS DISTAL FEMUR; P/N 25001040E, ELEOS MALE FEMALE MIDSECTION; P/N 25001140E, ELEOS MALE FEMALE MIDSECTION; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization;
Patient Age36 YR
Patient SexFemale
-
-