• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL REFLEXION¿ SPIRAL BI-DIRECTIONAL VARIABLE RADIUS CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL REFLEXION¿ SPIRAL BI-DIRECTIONAL VARIABLE RADIUS CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D402893
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 11/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
At the beginning of an atrial fibrillation procedure, when mapping was started with the tacticath se and reflexion spiral catheter, an acute pericardial effusion was noted on the ice image.The atrial fibrillation case was begun, transseptal was performed and left atrial mapping geometry commenced.The physician thought the geometry looked odd and asked to see the prior study from (b)(6) 2017.After reviewing that study he returned tableside to continue and noted an effusion on the ice image.The procedure was stopped and a pericardiocentesis was performed.There were multiple perforations on the left side of the heart.The cardiovascular surgeon in house was notified and the patient was eventually taken to have a presumed perforation repaired.The patient was extubated (b)(6) and is stable.There was no alleged issue with any abbott device.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.In addition, no log files were received.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported pericardial effusion remains unknown.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Additional Manufacturer Narrative
Corrected information: h6.Additional information: g1, g3, h2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFLEXION¿ SPIRAL BI-DIRECTIONAL VARIABLE RADIUS CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer (Section G)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15996391
Report Number2182269-2022-00065
Device Sequence Number20371233
Product Code DRF
UDI-Device Identifier05414734216654
UDI-Public05414734216654
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K072012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model NumberD402893
Device Catalogue NumberD402893
Device Lot Number8532047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer12/20/2022
12/01/2023
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date01/09/2023
12/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-