SMITH & NEPHEW, INC. PROBE,SAPHYRE II, 90DEG,3MM,SUCTION; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
| Model Number |
7210111 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopic shoulder surgery, the saphyre probe had an electrical leakage, there were sparks.The procedure was completed with a s+n back up device.There was no surgical delay and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was not received for evaluation.However, another device was received.A visual inspection revealed no manufacturing abnormalities.The electrode shows no wear or damage.The shaft has minor physical markings but no burn marks or arc damage.Product was out of the original packaging.No packaging returned.The cable of the device has been cut off.The suction line is still present.A functional evaluation could not be performed due to the cut cable.An analysis of the customer provided images found what appears to be the tip of the wand.It has plasma on the side, and looks like its heavily damaged in some way.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include a broken wire inside the wand cable, a damaged wire between the transition of the handle and the shaft, and damage to the shaft tip or housing.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|
|
|