• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PROBE,SAPHYRE II, 90DEG,3MM,SUCTION; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PROBE,SAPHYRE II, 90DEG,3MM,SUCTION; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 7210111
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopic shoulder surgery, the saphyre probe had an electrical leakage, there were sparks.The procedure was completed with a s+n back up device.There was no surgical delay and no further complications were reported.
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not received for evaluation.However, another device was received.A visual inspection revealed no manufacturing abnormalities.The electrode shows no wear or damage.The shaft has minor physical markings but no burn marks or arc damage.Product was out of the original packaging.No packaging returned.The cable of the device has been cut off.The suction line is still present.A functional evaluation could not be performed due to the cut cable.An analysis of the customer provided images found what appears to be the tip of the wand.It has plasma on the side, and looks like its heavily damaged in some way.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include a broken wire inside the wand cable, a damaged wire between the transition of the handle and the shaft, and damage to the shaft tip or housing.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROBE,SAPHYRE II, 90DEG,3MM,SUCTION
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15996471
Report Number3003604053-2022-00070
Device Sequence Number5922543
Product Code GEI
UDI-Device Identifier03596010481498
UDI-Public03596010481498
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K991218
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/10/2024
Device Model Number7210111
Device Catalogue Number7210111
Device Lot Number2071708
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2022
Supplement Date Received by Manufacturer03/05/2024
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date03/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-