Model Number C702
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Na.
 
Event or Problem Description
The initial reporter complained of a discrepant low result for 1 patient sample tested for tina-quant c-reactive protein iv (crp4) on a cobas 8000 c 702 module.The initial result was 0.00 mg/dl.The repeat result was 1.19 mg/dl.The questionable result was not reported outside of the laboratory.The crp4 reagent lot number was 653390.The expiration date was not provided.
 
Additional Manufacturer Narrative
The crp4 reagent lot number expiration date was 31-jul-2023.It was clarified that the repeat result was from a different c702 module."abnormal aspiration" and "sample short" errors were observed on the customer's alarm trace data from the day of the event.Based on the errors from the alarm trace data, the investigation determined the event was consistent with a preanalytical handling error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 C 702 MODULE
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15996628
Report Number1823260-2022-04047
Device Sequence Number12242944
Product Code JJE
UDI-Device Identifier04015630946846
UDI-Public04015630946846
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K100853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC702
Device Catalogue Number06473245001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
Patient SexMale
-
-