SYNTHES GMBH INSERT-HANDLE F/EXPERT TN+FN; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 03.010.045
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the procedure on (b)(6) 2022, the surgeon had an antegrade rafn fully inserted into the femur.He began to rotate the nail with the insertion handle.An audible pop was heard and it was evident that something was amiss when the proximal targeted interlocks missed through the aiming arm.Upon completion of the procedure, it was noticed that the tooth that connects the nail to the insertion handle had broken off, making the insertion handle rotate freely around the nail.The tooth remained in the proximal portion of the nail.The patient had a left femoral shaft fracture.There was a surgical delay of 8 minutes.Perfect circles were completed proximally, and the surgery was completed successfully.The fragments were retained in the patient.There was no clinical outcome experienced by the patient.No other medical intervention was necessary.This report involves one standard insertion handle.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The pin in the driving cap came out and the spring mechanism fell apart.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part: 03.010.045, lot: 1620767, manufacturing site: hägendorf, release to warehouse date: 19.Jan.2007.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Note: device history record review was acquired from product investigation (b)(4), as both impacted products share the same lot number.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the extrusion at the end of insert-handle f/expert tn+fn, p/n: broke off.The broken fragment was not returned for evaluation.The embedded device allegation can not be confirmed since no evidence to assess the condition was provided.A dimensional inspection was unable to be performed due to post-manufacturing damage.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the insert-handle f/expert tn+fn, p/n: would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: rev n current, l manufactured dimensional inspection: n/a.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSERT-HANDLE F/EXPERT TN+FN
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES HAEGENDORF
im bifang 6
haegendorf CO 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15996964
Report Number8030965-2022-11219
Device Sequence Number17584369
Product Code LXH
UDI-Device Identifier07611819760837
UDI-Public(01)07611819760837
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.010.045
Device Catalogue Number03.010.045
Device Lot Number1620767
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date01/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - NAILS: RAFN
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-