Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary as reported, an open-end flexi-tip ureteral catheter separated in the ureter during a retrograde pyleogram procedure.The separation was noticed after the stent was placed.The remaining portion of the flexi-tip was retrieved from the kidney with a stone extractor.The procedure was completed as normal.It was reported that the patient did not have any scarred or tortuous anatomy.The patient did not experience any adverse effects or require any additional procedures due to this occurrence.Investigation ¿ evaluation reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures of the device were conducted during the investigation.The device was not returned to cook for investigation.The device was reported to be disposed of after the procedure.A photo was provided of the complaint device which shows it was separated at the fifth ink mark from the tip.The catheter material appeared to be torn at the point of separation.The dhr records no non-conformances related to the reported failure mode.The flexi-tip bond strength test data for the lot was reviewed and all samples passed the test.A database search for complaints on the reported lot found no additional complaints reported from the field.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of relevant manufacturing documents was conducted.Cook has concluded that sufficient inspection activities are in place to assure functionality and device integrity prior to shipping.The instructions for use (ifu), includes the following precautions: precautions ¿ these devices are intended for use by physicians trained and experienced in urological techniques.Standard techniques should be employed.¿ avoid bending or kinking the catheter prior to placement.Doing so could damage the integrity of the catheter and result in patient injury.¿ if you encounter resistance while advancing or withdrawing the catheter, stop.Determine the cause of the resistance before proceeding.¿ do not withdraw catheter while deflected in cystoscope/endoscope.Based on the available information, cook has concluded a specific cause of the complaint cannot be established, it not possible to rule out inadvertent user/procedural issues (procedure complications and device handling before and during use are unknown).Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|