| Model Number |
R SERIES |
| Medical Device Problem Codes |
No Display/Image (1183); Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
Necrosis (1971)
|
| Date of Event |
11/26/2022
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate an 81-year-old male patient, the device powered on without display.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device and is also correcting information submitted on the initial medwatch report.Please reference sections b5 and h6 (device problem code).Evaluation results: the customer's report was observed during review of the device data logs.However, the device was put through extensive testing including bench handling, impedance testing, defibrillation cycling and environmental chamber testing without duplicating the report.An internal inspection found no discrepancies.The ac charger was replaced as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate an 81-year-old male patient, the device would not power up.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
|
| |
|