| Model Number |
SCX-375 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Failure of Implant (1924)
|
| Date of Event |
07/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and capsular contracture baker grade unknown.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported rupture and capsular contracture baker grade unknown.Device has been explanted.This record is for the right side.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right-side rupture and capsular contracture, baker grade iv and hole non- specific.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, right-side rupture.And capsular contracture, baker grade iv.And hole non- specific.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, a broken assessed, as a surgical impact opening (shell thickness whiting to spec).Capsular contracture: unable to observe, as it is a medical event.Not related to the device.Additional observations: stress mark observed, on the device.No further actions are required, as the device was implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d4, d9, h3, h4, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right-side rupture and capsular contracture, baker grade iv and hole non- specific.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|