| Model Number |
60005 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Eye / Vision Problem (4471)
|
| Date of Event |
11/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The site reported that a light adjustable lens (lal, sn (b)(4), 23.5d) was explanted due to it being delivered backwards into the eye.Rxsight's first awareness of the event was on 11/18/2022.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.The original lal (sn(b)(4), 23.5d) was implanted backwards.The surgeon opted to explant the lens during the initial surgery and replace it with another lal of the same power (23.5d).The original incision was enlarged from 2.8mm to 6.0mm to explant the lens and sutures were used to close the wound, which was a deviation from the surgical plan.The replacement lal was successfully delivered into the eye without further complications.The patient is doing well after surgery.The lal is centered in position.Manufacturer reference #: (b)(4).
|
| |
|