• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RXSIGHT, INC. LIGHT ADJUSTABLE LENS (LAL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RXSIGHT, INC. LIGHT ADJUSTABLE LENS (LAL) Back to Search Results
Model Number 60005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Eye / Vision Problem (4471)
Date of Event 11/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The site reported that a light adjustable lens (lal, sn (b)(4), 23.5d) was explanted due to it being delivered backwards into the eye.Rxsight's first awareness of the event was on 11/18/2022.
 
Additional Manufacturer Narrative
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.The original lal (sn(b)(4), 23.5d) was implanted backwards.The surgeon opted to explant the lens during the initial surgery and replace it with another lal of the same power (23.5d).The original incision was enlarged from 2.8mm to 6.0mm to explant the lens and sutures were used to close the wound, which was a deviation from the surgical plan.The replacement lal was successfully delivered into the eye without further complications.The patient is doing well after surgery.The lal is centered in position.Manufacturer reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT ADJUSTABLE LENS (LAL)
Common Device Name
LIGHT ADJUSTABLE LENS (LAL)
Manufacturer (Section D)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
jim zhang
100 columbia
aliso viejo, CA 92656
9492736163
MDR Report Key15999693
Report Number3012712027-2022-00069
Device Sequence Number12210758
Product Code PZK
UDI-Device Identifier00818806020746
UDI-Public01008188060207461122070117250630
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model Number60005
Device Catalogue Number60005
Device Lot NumberL07-004071
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2022
Initial Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-