| Model Number |
392544 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bd insyte autoguard bc pro straight needle would not retract.The following information was provided by the initial reporter: i have received more of the affected lot and a couple of the products which have been unsuccessfully safetied.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Date of event: unknown.The date received by manufacturer has been used for this field.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 16-dec-2022.H6: investigation summary: our quality engineer inspected the samples submitted for evaluation.Bd received five used 18g insyte autoguard bc pro units with no product documentation to indicate reference or lot number.A gross visual inspection found that unit (b)(4) had a spring that was snapped into two pieces.The remaining units were tested for retraction, and all retracted as intended.A damaged spring as seen in unit (b)(4) would likely cause a partial retraction or no retraction at all.Your reported issue was confirmed.A damaged spring may occur during manufacturing due to incorrect equipment settings on the spring feeder (feeder that loads the springs into the grips) or due to a machine misalignment as the needle hub is loaded into the grip.A 100% spring force check is performed in line and operators perform functional testing for needle retraction per the quality plan to mitigate the occurrence of this defect.In addition, a device history record review showed no non-conformances associated with this issue during the production of this batch.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd insyte autoguard bc pro straight needle would not retract.The following information was provided by the initial reporter: i have received more of the affected lot and a couple of the products which have been unsuccessfully safetied.
|
| |
|