| Model Number |
0475000100 |
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a thermal event for burning smell.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: burning smell probable root cause: console fuses fail to break circuit due to excessive mains current leading to fire isolation power supply failure use error: console is sterilized or disinfected use error: console cleaned with incompatible products the reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a thermal event for burning smell.
|
| |
|
Search Alerts/Recalls
|