• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORPORATION COBLATOR II CONTROLLER (240V); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORPORATION COBLATOR II CONTROLLER (240V); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number EC8001-01
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an ent procedure, when plugged in the coblator ii system did not turn on.A new smith and nephew back-up device was immediately sent which caused a 2 hours delay until the surgeon could finish the procedure, and the patient was under anesthesia all this time.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H2: additional information on a2 and a3.
 
Event or Problem Description
It was reported that, during an ent procedure, when plugged in the coblator ii system, the pius burned and the device didn´t turn on.A new smith and nephew back-up device was immediately sent which caused a 2 hours delay until the surgeon could finish the procedure.The doctor had to wake the patient up and put him to sleep again two hours later when the replacement device arrived.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h2: additional information on b3 and b5.
 
Additional Manufacturer Narrative
H10 h3,h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Visual inspection observed the warranty seal is broken, the screws have signs of being removed previously, and the fuse housing is cracked inside from being accessed.The fuses in the fuse housing are installed incorrectly.Functional evaluation revealed the unit has no power when the switch is pressed.Checking the fuses in the fuse housing, they were installed incorrectly by a third party, and one is blown.Replacing and installing the fuses correctly resolved the issue.The complaint was confirmed and the root cause is associated with a component failure.Factors which could have contributed to the identified malfunction, include a blown fuse.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBLATOR II CONTROLLER (240V)
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16000192
Report Number3006524618-2022-00509
Device Sequence Number17544584
Product Code GEI
UDI-Device Identifier00817470002164
UDI-Public00817470002164
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K030108
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEC8001-01
Device Catalogue Number13546-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2022
Supplement Date Received by Manufacturer12/20/2022
12/28/2022
07/17/2023
Initial Report FDA Received Date12/16/2022
Supplement Report FDA Received Date12/28/2022
12/30/2022
07/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age12 YR
Patient SexMale
-
-