| Catalog Number |
2000-9061 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that a warehouse found a twisted tip on a polaris 5.5 plug driver upon return from a hospital.The facility did not provide any patient or surgical information with the returned device.
|
| |
|
Event or Problem Description
|
|
It was reported that a warehouse found a twisted tip on a polaris 5.5 plug driver upon return from a hospital.The facility did not provide any patient or surgical information with the returned device.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection revealed that the driver tip was twisted.Root cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.It could also be attributed to applying off-axis forces on the driver during use.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|