• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM; CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM; CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012582-19
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device problem code (b)(4) - use after expiration.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.It was reported that the graftmaster was used past the expiration date.A review of the graftmaster label attached to the lot history record for this lot was conducted indicating a manufacturing date of (b)(6) 2020 with an expiration date (use by date) of 31(b)(6) 2022.The product was used after expiration as it was reported the procedure occurred on (b)(6) 2022.The expiration date of the product is important for sterility, efficacy, and performance of the device.It should be noted that the graftmaster rapid exchange (rx), coronary stent graft system, domestic instructions for use, states: note the product use by date specified on the package.It is likely the ifu deviation contributed to the reported event.The investigation determined the reported device expiration issue appears to be related to use error.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a perforation in the saphenous vein graft to right coronary artery (svg to rca).The 4.0x19mm graftmaster covered stent was successfully implanted at a pressure of 16 atmospheres (atms), however, the device was expired and it was noted after the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Common Device Name
CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key16001526
Report Number2024168-2022-12791
Device Sequence Number5784206
Product Code MAF
UDI-Device Identifier08717648176456
UDI-Public08717648176456
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number1012582-19
Device Catalogue Number1012582-19
Device Lot Number0112341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2022
Initial Report FDA Received Date12/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age84 YR
Patient SexMale
Patient Weight82 KG
-
-