| Model Number |
254651 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).The lot number was unknown.The lot number was unknown; therefore, the expiration date, manufacturing site name and device manufacture date were unknown.Reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported by the sales rep in italy that preoperatively to an unknown surgery on (b)(6) 2022, it was observed that the biointrafix trial device was worn out and maybe broken.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was not returned after multiple attempts for device return, therefore unavailable for a physical evaluation.Should the device ever be received back in the future, this complaint file will be reopened at that time and an evaluation will be performed and documented.No lot numbers were supplied which precludes conducting a dhr review or a lot of specific search in the complaints handling system.As part of depuy mitek¿s quality process all devices are manufactured, inspected, and released to approved specifications.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a ollow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary - the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Visual observations confirms that the device was received broken and separated in two pieces at the weld point near the connection handle-shaft.In addition, is dull and the device appears worn.The complaint can be confirmed.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.This type of failure is typically observed after the device has been reprocessed over the course of many years therefore causing the weld integrity to degrade.However, given the information provided we cannot discern a definitive root cause for the reported failure.As per ifu, inspect for damage prior to use; and replace a damaged, worn or bent instrument.Also, the end of useful instrument life is generally determined by wear or damage from handling or surgical use.Inspect instruments between uses to verify proper functioning.Finally, it is necessary to follow the instructions and warnings of any cleaning and disinfection agents and equipment used; also, the recommendation of the proper condition during the sterilization and maintenance to avoid any damage.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Search Alerts/Recalls
|
|