MAUDE Adverse Event Report: WILLIAM A. COOK AUSTRALIA, PTY LTD ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM A. COOK AUSTRALIA, PTY LTD ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number G32539
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Clinical Endoleak (4843)
Type of Reportable Event Serious Injury
Event or Problem Description
A patient of undisclosed gender and age underwent an abdominal aortic aneurysm repair in (b)(6) 2018 in which a zenith fenestrated graft was implanted.The patient was lost to follow- up for quite some time, however, came back recently for a follow up.Per recent scan the physician detected a possible type 1a endoleak.The physician is planning a follow up procedure but not scheduled as of yet.For the follow up procedure the physician will be placing a non-cook manufactured stent.
 
Event or Problem Description
A patient of undisclosed gender and age underwent an abdominal aortic aneurysm repair in (b)(6)2018 in which a zenith fenestrated graft was implanted.The patient was lost to follow- up for quite some time, however, came back recently for a follow up.Per recent scan the physician detected a possible type 1a endoleak.The physician is planning a follow up procedure but not scheduled as of yet.For the follow up procedure the physician will be placing a non-cook manufactured stent.
 
Additional Manufacturer Narrative
No device was returned for evaluation as it remains implanted in the patient.Medical imaging was received to assist with the investigation.The imaging was reviewed by the medical director who noted that the device was initially implanted in (b)(6) 2018 and the patient was lost to follow up but returned for imaging on 15-nov-2022.The medical director commented the following on the ct imaging available: ¿the ct shows a large endoleak almost completely filling the aaa sac in front of the endograft.¿ the medical director further commented that it appears to be a type iii endoleak rather than initially reported type ia endoleak: ¿¿and arising mainly from the right renal fenestration, where the right renal bridging stent appears to have pulled out.There will be forward blood flow out through the open right renal fenestration, plus backflow from the right renal bridging stent.There is a lot of radio-opaque artifact on the ct scans in the vicinity of the right renal fenestration, which could be consistent with multiple coils that may have been deployed to try and stop the endoleak, although there is no mention of this in the report.¿ due to the mass of artifact the medical director noted it was impossible to accurately determine the site of the endoleak.It was stated the there was no visible evidence of flow getting outside proximally of the neck of the graft.The medical director commented that: ¿the most likely cause for this would be inadequate flaring of the right renal bridging stent, then gradual withdrawal of that stent from the fenestration of the intervening period.I think this is a failure due to a usage problem." additional information was received from the physician via the district manager.The physician thinks the endoleak was due to the right renal artery stent not extending into the aorta and it appeared that it could not have been flared much to further help with the seal.It was discussed that the endoleak would be treated by re-lining the right renal artery with a longer stent to extend further and flare more in the aorta, re-ballooning all junctures and potentially stent the superior mesenteric artery if warranted.The work order was reviewed for lot ac94549 and appears complete and the quality control inspection was verified to ensure the device passed inspection.There were no non-conformances raised or temporary deviations in place at the time of manufacture.It was concluded that the device was manufactured to specification.The instructions for use supplied with the device was reviewed and contains information that the long-term performance of fenestrated endovascular grafts, including the stents placed in fenestrations/scallops, have not yet been established.Patients are advices that endovascular treatment requires life-long, regular follow-up to assess their health and the performance of their endovascular graft.It is listed that inaccurate placement and/or incomplete sealing of the device within the vessel may result in increased risk of endoleak, migration or inadvertent occlusion of the renal or internal iliac arteries.Based on the available information the investigation concluded that the most likely root cause is inadequate flaring of visceral stent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY
Common Device Name
MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
WILLIAM A. COOK AUSTRALIA, PTY LTD
95 brandl street
eight mile plains
brisbane
AS 
Manufacturer (Section G)
COOK AUSTRALIA
brisbane technology park
95 brandl street
eight mile plains, qld 4113
AS   4113
Manufacturer Contact
majbrit frosig
brisbane technology park
95 brandl street
eight mile plains, qld 4113
AS   4113
738411188
MDR Report Key16001996
Report Number9680654-2022-00043
Device Sequence Number20502948
Product Code MIH
UDI-Device Identifier10827002325392
UDI-Public(01)10827002325392(17)180211(10)AC949549
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P020018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/11/2018
Device Model NumberG32539
Device Catalogue NumberZFEN-P-2-32-109-R
Device Lot NumberAC949549
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date12/19/2022
Supplement Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/11/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-