| Model Number |
G47496 |
| Medical Device Problem Codes |
Malposition of Device (2616); Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturers ref#: (b)(4).Pma/510(k): p180001.Investigation is still in progress.This report is required by the fda under 21cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Description of event according to initial reporter: the tevar covered pieces were successfully deployed to cover entry tears to approx.2cm above celiac artery.Extended distally with zdes-46-184-us.The physician was adviced to use caution advancing any wires through the bare stent as they could go through the suture loops used to hold the stent together, but he said he advanced wires quickly and said that there was no way to tell.As it was explained to the fellow and mdt rep why caution was needed, the physician advanced the berenstein catheter over the wire at which point the distal 2 stents of the zdes migrated up.The doctor used a molding balloon (from another mfg) that was gently inflated inside the zdes to pull the stent back down into place.Patient outcome: the stent was in a reasonable position after using the molding balloon and stayed in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer ref# (b)(4).Summary of investigational findings: a 67-year-old female patient underwent tevar to treat an aortic dissection with malperfusion.The tevar covered pieces were successfully deployed to cover entry tears to approx.2cm above celiac artery.Extended distally with zdes-46-185-us.The sales representative advised the doctor to use caution advancing any wires through the bare stent as they could go through the suture loops used to hold the stent together, but the physician said he advanced wires quickly and said that there was no way to tell.As the sales representative was explaining to the fellow and mdt rep why caution was needed the doctor advanced the berenstein catheter over the wire at which point the distal 2 stents of the zdes migrated up (cephalad).The physician used a molding balloon (mdt reliant) that was gently inflated inside the zdes to pull the stent back down into place.No adverse effects to the patient are reported.Review of the device history record gave no indication of the device being produced out of specification.Procedural angiogram is provided, and an imaging review is performed by an imaging expert.Per the imaging review the complaint event is confirmed.The following is the imaging reviewers¿ findings ¿a zdes (46 x 185 mm) dissection stent is then placed from the from the distal thoracic graft to the distal abdominal aorta and appears fairly well expanded.Advancement of a catheter from the right iliac access directly pushes the distal zdes stent upward about 2 cm, resulting in partial stent overlap of the distal 5 stent segments.¿ and ¿a large caliber balloon (medtronic) is inflated within the distal zdes stent and used to pull the zdes back down with only mild persistent overlap of the distal 4th and 5th stent segments.¿ based on these findings the imaging reviewers¿ impressions are ¿after placement of the zdes stent, a catheter is advance from the right iliac access and inadvertently pushes the distal stent up about 2 cm.A large balloon is inflated within the distal zdes and used to successfully pull the stent back down to near its original position.The distal zdes stent appears well expanded prior to the migration.It is unclear if the wire/catheter were inadvertently passed through the interstices of the distal zdes stent, but the catheter is advanced quickly across this segment and clearly pushes the zdes stent up.The catheter advancement stops only after the stent has already been pushed up about 2 cm.¿ per the instruction for use "exercise caution when manipulating a wire guide through an in-situ zenith dissection endovascular stent; the wire guide may become entangled with the stent." based on the provided information and the imaging review it is unclear if the wire/catheter were inadvertently passed through the interstices of the distal zdes stent, but the 2 stents of the zdes migrated up as they were being pushed during advancement of the berenstein catheter.In addition it is noted that the zdes is used in combination with a medtronic endograft which is outside the recommendations for this device.No evidence to suggest that the device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|