• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH BLUNT TIP SCREW, 4X34MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH BLUNT TIP SCREW, 4X34MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number N/A
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent initial right shoulder surgery.Approximately 2 months later, revision surgery was performed due to screw migration.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).Proximal humerus, right, 7x160mm; item# 47249616007 lot# 3091684.Blunt tip screw, 4x38mm; item# 47248603840 lot# 3082107.Blunt tip screw, 4x42mm; item# 47248604240 lot# 3081945.Blunt tip screw, 4x50mm; item# 47248605040 lot# 3078232.Cortical bone screw, 4x24mm; item# 47248612440 lot# 3091458.Cortical bone screw, 4x24mm; item# 47248612440 lot# 3091460.Proximal humerus nail cap, 0mm; item# 47248801000 lot# 3091386.Report source: foreign: japan.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350 -2022 -00635, 0009613350 - 2022 -00637, 0009613350 -2022 -00638.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h1, h2, h6, h10.As no products were returned, visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Devices are used for treatment.No medical reports were provided.Based on the investigation it could be assumed that further possible contributing factors to the migration of the screw might be multifactorial related to either patient condition, behavior or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the locking mechanism of the nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the nail proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.Therefore, no design changes were conducted.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUNT TIP SCREW, 4X34MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key16003190
Report Number0009613350-2022-00636
Device Sequence Number6738152
Product Code HSB
UDI-Device Identifier00889024505414
UDI-Public(01)00889024505414(17)260714(10)3077767
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200814
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/14/2026
Device Model NumberN/A
Device Catalogue Number47248603440
Device Lot Number3077767
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/19/2022
Supplement Report FDA Received Date01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexPrefer Not To Disclose
-
-