| Model Number |
E601 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of a questionable cea elecsys result from cobas 6000 e601 module serial number (b)(4).The initial result was 0.97 ng/ml and was reported outside of the laboratory.The qc performed afterward was not acceptable.The customer recalibrated the assay and ran qc which was acceptable.The customer then repeated the sample.The repeat result was 1.54 ng/ml and was deemed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined the issue was resolved by the customer recalibrating the assay.
|
| |
|