(b)(4).Batch #: unknown.Additional information was requested and the following was obtained: what was the location of the tumor? right lower pole.What is the relationship of the tumor to the urinary collecting system? 3 cm mass.Mostly exophytic, but extending to an underlying calyx.How many probes were used during the ablation? 3.What were the ablation settings, such as number of deliver sessions, length of deliver, and power level? 65 watts, 5 min for all 3 probes.How was the uretic stricture and urinoma detected? 6 month follow-up mri.How was the uretic stricture and urinoma managed for the patient? right ureteral stent placement.What is the patient's current status? outpatient.In the surgeon¿s opinion what was the cause of the extended ablation zone? compression of tissues due to lateral position, extension of ablation zone more than 5 mm from the tips of the probes.An analysis of the product could not be performed since a physical sample was not received for evaluation.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|