• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RENEGADE; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RENEGADE; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 29417
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the catheter was broken.A 105/3.0/2.5/.021/20 renegade microcatheter was selected for use.Upon unpacking, it was noted that the catheter was broken.The procedure was completed with another of the same device.No complications reported and patient was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.The device was visually and microscopically inspected for damage.The device showed damage in the form of a stretched and fractured shaft located 7.3cm from the hub.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The device was confirmed for stretching and a fracture.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that the catheter was broken.A 105/3.0/2.5/.021/20 renegade microcatheter was selected for use.Upon unpacking, it was noted that the catheter was broken.The procedure was completed with another of the same device.No complications reported and patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16005181
Report Number2124215-2022-53361
Device Sequence Number5783839
Product Code DQO
UDI-Device Identifier08714729375395
UDI-Public08714729375395
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K020012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/04/2025
Device Model Number29417
Device Catalogue Number29417
Device Lot Number0028807812
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/28/2022
Supplement Date Received by Manufacturer01/18/2023
Initial Report FDA Received Date12/19/2022
Supplement Report FDA Received Date02/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexFemale
-
-