MAUDE Adverse Event Report: ZIMMER CAS CAS FIX PIN 3.2D X 150MM STR; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER CAS CAS FIX PIN 3.2D X 150MM STR; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number N/A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that total knee arthroplasty was performed with rosa knee.When fixed navi-tracker to the patient femoral bone with the pins, surgeon had a hard time to insert them because the patient bone quality was very good.When the surgeon removed them to re-insert, one of them broke at the tip.Subsequently, another pin was used to complete the procedure.Postoperatively, the broken tip was seen on x-ray.No revision is planned.No additional information.
 
Additional Manufacturer Narrative
(b)(4).Report source: foreign: country: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Physical and technical evaluation was performed.The reported fractured tip of the pin was confirmed.Additionally two pictures provided also display the broken pin tip.Review of the dhr for p/n 20-8000-00-10, lot # 65443479 (cas fix pin 3.2d x 150mm str) did not identify any related deviations or anomalies which would have caused or contributed to the reported event.The root cause of the broken pin could not be determined precisely with the information available.Possible contributing factors are excessive force and/or off-angle application during insertion/removal.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAS FIX PIN 3.2D X 150MM STR
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
ZIMMER CAS
75 queen str.
suite 3300
montreal, quebec H3C 2 N6
CA  H3C 2N6
Manufacturer (Section G)
ZIMMER CAS
75 queen str.
suite 3300
montreal, quebec H3C 2 N6
CA   H3C 2N6
Manufacturer Contact
kristen lashley
56 e. bell dr.
warsaw, IN 46582
9016334069
MDR Report Key16005571
Report Number0009617840-2022-00030
Device Sequence Number6577890
Product Code OLO
UDI-Device Identifier00889024563629
UDI-Public(01)00889024563629(17)270424(10)65443479
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K210121
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number20-8000-000-10
Device Lot Number65443479
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2022
Supplement Date Received by Manufacturer04/12/2023
Initial Report FDA Received Date12/19/2022
Supplement Report FDA Received Date06/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-