| Model Number |
383336 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medical device lot #: 1287217 was reported, however, this is not a lot # manufactured for the reported catalog #.Medical device expiration date: unknown.Device manufacture date: unknown.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd saf-t-intima¿ closed iv catheter system the needle broke.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the customer that they had issues with the broken saf t intima.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary a device history review was conducted for lot number 1287217.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.See h10.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd saf-t-intima¿ closed iv catheter system the needle broke.There was no report of patient impact.The following information was provided by the initial reporter: it was reported by the customer that the customer that they had issues with broken saf t intima.
|
| |
|