| Model Number |
ECS25B |
| Medical Device Problem Code |
Ejection Problem (4009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a laparoscopic total hysterectomy and salpingo-oophorectomy, greater omentum resection and sigmoidectomy, the anvil head could not be disassembled from the device before use.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 01/18/2023.D4: batch # 975a85.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ecs25b device arrived with no apparent damage and along with a tyvek.The breakaway washer was present, uncut and the device was fully loaded with staples.Further analysis of the device showed that the trocar was damaged.However, the anvil was installed onto the trocar without any difficulties.In addition, the device was tested with the returned washer for functionality; the device formed all the staples, as well as completely cut the test media and the breakaway washer without incident.The staple line was complete and the staples meet the staple form release criteria.This defect has been correlated to a manufacturing process.A manufacturing record evaluation was performed for the finished device batch number 975a85, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|