• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ANGIOJET ULTRA 5000A; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ANGIOJET ULTRA 5000A; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 45043
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a procedure was cancelled.An angiojet console was selected for thrombectomy procedure.However, during the use of the unit, it was noticed that there was liquid leaking and suspected a bad drawer assembly.The procedure was not completed due to this event.No further complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was not returned for evaluation.However, an in field work order was reviewed as part of this investigation.The unit was observed leaking and suspected to be a bad drawer assembly.A boston scientific field service engineer (fse) visited the site on 20feb2023.The replacement part was delivered to the customer site to install the component.After install, the new component would not calibrate, and the new component was deemed faulty and would not work.On 22feb2023 the fse returned to the customer site and installed a new drawer component.The console was verified to be functional, and unit is ready for use.
 
Event or Problem Description
It was reported that a procedure was cancelled.An angiojet console was selected for thrombectomy procedure.However, during the use of the unit, it was noticed that there was liquid leaking and suspected a bad drawer assembly.The procedure was not completed due to this event.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET ULTRA 5000A
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
150 baytech dr.
san jose CA 95134
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16007000
Report Number2124215-2022-52490
Device Sequence Number5620037
Product Code MCX
UDI-Device Identifier08714729890607
UDI-Public08714729890607
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
P980037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number45043
Device Catalogue Number45043
Device Lot NumberU4485
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer02/21/2023
Initial Report FDA Received Date12/19/2022
Supplement Report FDA Received Date03/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-