| Catalog Number |
300613 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd plastipak¿ syringe stopper separated from the plunger during product sampling.The following information was provided by the initial reporter, translated from french: "the rubber part came off during a product sampling.Occurred in our radiology unit.The rubber part came off during a product sampling.".
|
| |
|
Event or Problem Description
|
|
It was reported that the bd plastipak¿ syringe stopper separated from the plunger during product sampling.The following information was provided by the initial reporter, translated from french: "the rubber part came off during a product sampling.Occurred in our radiology unit.The rubber part came off during a product sampling.".
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 02-jan-2023.H6: investigation summary: one sample was provided to our quality team for investigation.The product was visually inspected and the stopper was verified to be incorrectly assembled.A device history review was performed for lot 2207134, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.
|
| |
|
Search Alerts/Recalls
|