W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number CXA230005
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Obstruction/Occlusion (2422)
Date of Event 11/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, the patient underwent endovascular treatment of a distal stent graft-induced new entry tear caused by a non-gore manufactured stent graft using two gore® excluder® conformable aortic extender endoprostheses.During the procedure, two aortic extender devices were deployed, and the distal device was placed just above the lumbar artery that was to be preserved.However, the two aortic extender devices reportedly migrated approximately 5mm distally while ballooning, and the lumbar artery was unintentionally covered by the distal aortic extender device.The physician elected to monitor the vessel coverage, and the procedure was completed.The patient tolerated the procedure.It was reported the device migrated due to the balloon being pushed by the blood flow.
 
Additional Manufacturer Narrative
(b)(4).It should be noted that, per the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and / or require intervention include, but are not limited to, improper endoprosthesis component placement, occlusion of device or native vessel.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
damon jackson
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key16010341
Report Number3007284313-2022-02273
Device Sequence Number6570401
Product Code MIH
UDI-Device Identifier00733132650934
UDI-Public00733132650934
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/24/2024
Device Model NumberCXA230005
Device Catalogue NumberCXA230005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2022
Initial Report FDA Received Date12/19/2022
Date Device Manufactured08/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age33 YR
Patient SexMale
-
-