Model Number EB-1970UK
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We checked the returned unit and confirmed that the ccd module distorted image.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the ccd driver pcb fluid damage, the lg cable connector fluid damage, the light guide fiber bundle (lcb) broken, the insertion flexible tube (ift) buckled, the light guide cable buckled, the lg cable connector corroded, the balloon feeder socket deformed, the remote control buttons cut, the distal body worn out, the warning/caution label worn out, the us connector body defective, the operation channel resistance, and the us probe eus image no scanline; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586 (image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Video image failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
LINEAR ULTRASOUND VIDEOBRONCHOSCOPE 2.0C
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key16011559
Report Number9610877-2022-61283
Device Sequence Number7303022
Product Code EOQ
UDI-Device Identifier04961333148226
UDI-Public04961333148226
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K131946
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEB-1970UK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2022
Initial Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-