Model Number N/A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Report source foreign: japan.Proximal humerus, right, 9x160mm #item 47249616009 #lot 3083730; blunt tip screw, 4x38mm #item 47248603840 #lot 3081898; blunt tip screw, 4x40mm #item 47248604040 #lot 3086986; ann ph nail rt 9x160mm #item 47-2496-160-09 #lot 3083730; ann cort bone screw 4 x 24mm #item 47-2486-124-40 #lot 3091458; ann washer small #item 47-2488-000-04 #lot 3081906; affixus ph nl cap 0mm #item 47-2488-010-00 #lot 3091377.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2022-00640; 0009613350-2022-00641; 0009613350-2022-00642.
 
Event or Problem Description
It was reported that approximately two (2) months after implantation, during a routine post-operative checkup, it was found out that the leg screw had backed out from the proper position.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and corrected information.As no products were returned, visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Devices are used for treatment.The product combination was approved by zimmer biomet.The surgical technique explains that the locking of the corelock is done using the corelock driver with torque limiting handle."turn slowly clockwise to tighten and engage the corelock mechanism until a click is felt from the torque limiting handle." no medical reports were provided.Based on the investigation it could be assumed that further possible contributing factors to the migration of the screw might be multifactorial related to either patient condition, behavior or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism of the affixus nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the affixus® natural nail® proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.Therefore, no design changes were conducted.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLUNT TIP SCREW, 4X42MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key16012194
Report Number0009613350-2022-00643
Device Sequence Number12618971
Product Code HSB
UDI-Device Identifier00889024505452
UDI-Public(01)00889024505452(17)260922(10)3082110
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181827
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/22/2026
Device Model NumberN/A
Device Catalogue Number47-2486-042-40
Device Lot Number3082110
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer01/20/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;