Model Number 383336
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since no photos or samples displaying the reported condition of needle clogged / blocked were available for examination, we were unable to fully investigate this incident.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported while using bd saf-t-intima¿ closed iv catheter system the needles were clogged.There was no report of patient impact.The following information was provided by the initial reporter, translated from chinese to english: it was found that the needle core was blocked and the fluid could not be discharged, and then the patient was given a replacement the indwelling needle did not cause harm to the patient.
 
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
Becton dickinson and company's (bd) medication delivery solutions (mds) business unit will discontinue malfunction mdr reporting for certain device failures that have not caused or contributed to deaths or serious injuries in the past two years, and where the likelihood of a death or serious injury as a result of these malfunctions is remote.This decision follows fda guidelines (medical device reporting for manufacturers guidance for industry and food and drug administration staff issued nov 8, 2016, reference section 2.15).Bd notified fda of this decision on mar 3, 2025.Fda has reviewed and responded to bd¿s notification (reference fda document # (b)(4)) on march 7, 2025.This documentation is available in bd document management system (sap) as dir 10000690801.This supplemental mdr is being filed to document that the below device failure will no longer be considered a reportable malfunction mdr for the product family below: product family: saf-t-intima.Device failure: needle clogged / blocked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAF-T-INTIMA¿ CLOSED IV CATHETER SYSTEM
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16013121
Report Number9610847-2022-00488
Device Sequence Number6324418
Product Code FOZ
UDI-Device Identifier30382903833369
UDI-Public(01)30382903833369
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K923702
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2025
Device Model Number383336
Device Catalogue Number383336
Device Lot Number1117637
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2022
Supplement Date Received by Manufacturer05/02/2025
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date05/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-