Model Number 6570-0-036
Medical Device Problem Codes Corroded (1131); Degraded (1153); Naturally Worn (2988)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
 
Event or Problem Description
As reported: "head & liner were exchanged during an irrigation and debridement meant for infection.Stripe wear present on ceramic & liner.Surgeon reported corrosion present on neck/trunnion of femoral stem".
 
Additional Manufacturer Narrative
Reported event: an event regarding infection and wear involving a ceramic head was reported.The event was not confirmed.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device indicated that black debris is present on the inner head tapered surface where the stem trunnion would lock.The appearance of the striations is consistent with material transfer from the stem/biological material.No obvious wear on the ceramic head was observed.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot or sterile lot referenced.Conclusions: it was reported that the patient was revised due to infection and that wear of the ceramic head / corrosion at the stem-head interface was observed intraoperatively.Visual inspection of the returned device indicated that black debris is present on the inner head tapered surface where the stem trunnion would lock.The appearance of the striations is consistent with material transfer from the stem/biological material.No obvious wear on the ceramic head was observed.All stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance with applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
As reported: "head & liner were exchanged during an irrigation and debridement meant for infection.Stripe wear present on ceramic & liner.Surgeon reported corrosion present on neck/trunnion of femoral stem".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA V-40 CERAMIC HEAD 36/-5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood WD6 3-SJ
UK   WD6 3SJ
2082386500
MDR Report Key16013359
Report Number0002249697-2022-01858
Device Sequence Number20458759
Product Code JDI
UDI-Device Identifier04546540608482
UDI-Public04546540608482
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K052718
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model Number6570-0-036
Device Catalogue Number6570-0-036
Device Lot Number53800603
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/25/2022
Supplement Date Received by Manufacturer02/01/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date02/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age61 YR
Patient SexMale
Patient Weight109 KG
-
-