| Model Number |
6570-0-036 |
| Medical Device Problem Codes |
Corroded (1131); Degraded (1153); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
11/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
|
| |
|
Event or Problem Description
|
|
As reported: "head & liner were exchanged during an irrigation and debridement meant for infection.Stripe wear present on ceramic & liner.Surgeon reported corrosion present on neck/trunnion of femoral stem".
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding infection and wear involving a ceramic head was reported.The event was not confirmed.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device indicated that black debris is present on the inner head tapered surface where the stem trunnion would lock.The appearance of the striations is consistent with material transfer from the stem/biological material.No obvious wear on the ceramic head was observed.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot or sterile lot referenced.Conclusions: it was reported that the patient was revised due to infection and that wear of the ceramic head / corrosion at the stem-head interface was observed intraoperatively.Visual inspection of the returned device indicated that black debris is present on the inner head tapered surface where the stem trunnion would lock.The appearance of the striations is consistent with material transfer from the stem/biological material.No obvious wear on the ceramic head was observed.All stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance with applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
As reported: "head & liner were exchanged during an irrigation and debridement meant for infection.Stripe wear present on ceramic & liner.Surgeon reported corrosion present on neck/trunnion of femoral stem".
|
| |
|
Search Alerts/Recalls
|