| Model Number |
NSLG2S35 |
| Medical Device Problem Codes |
Device Alarm System (1012); Failure to Read Input Signal (1581); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # x95e9c.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nslg2s35 device was received with the top of the i-blade fractured and slightly lifted.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.Therefore, we were unable to investigate further the alert screen issues reported.The jaws were found attached to the instrument.In addition, no pieces were found missing under microscope inspection.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.A probable cause of the damage to the i-blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during a oophorectomy the doctor uses nslg2s35 during the procedure.The doctor occupies the clamp without any problem at the beginning of surgeries and during the procedure it began to mark that the clamp was opened, later it tells me to change the device and the nurse comments that the blade came off.Changed the clamp to continue with the process.There were no patient consequences.
|
| |
|
Search Alerts/Recalls
|