• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION HEDGEHOG; ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION HEDGEHOG; ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE Back to Search Results
Model Number SBD-228-50
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The exact date of the event is unknown.The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of the event.The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that a hedgehog double-end brush was used on an unknown date.During scope cleaning, the brush part came off.There was no patient involvement with this event.
 
Additional Manufacturer Narrative
Block b3: the exact date of the event is unknown.The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of the event.Block d4 and h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: device code a0401 captures the reportable event of brush break.Block h10: investigation results: the returned hedgehog double-end brush was analyzed.Visual evaluation found that both small brush heads were missing bristles.No other problems were noted.The reported event was confirmed.Product analysis identified that both small brush heads were missing bristles.It is possible for bristles to become dislodged if the brush is twisted or torqued while inside the scope, opposite the direction in which the bristles are twisted onto the shaft during assembly.Breaking of bristles can also occur if the brush is forced into the scope.According to the instructions for use (ifu) excessive twisting or torquing is not recommended.Based on all available information and analysis of the returned device, the most probable cause is unintended use error caused or contributed to event, indicating that the interaction between the user and device caused or contributed to the event.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a hedgehog double-end brush was used on an unknown date.During scope cleaning, the brush part came off.There was no patient involvement with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEDGEHOG
Common Device Name
ACCESSORIES, CLEANING BRUSHES, FOR ENDOSCOPE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
SUZHOU TEMEIRUL MEDICAL PRODUCTS CO
no 699 pinan road puzhuang
linhu to
suzhou, 100 21510 5
CH   215105
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16013552
Report Number3005099803-2022-07582
Device Sequence Number7334638
Product Code MNL
UDI-Device Identifier00816849010083
UDI-Public00816849010083
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSBD-228-50
Device Catalogue Number60252
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer01/31/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date02/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-