|
SMITH & NEPHEW, INC. TRIGEN LOW PROFILE SCREW 5.0MM X 42.5MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
71648240 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Fall (1848); Failure of Implant (1924); Arthralgia (2355); Implant Pain (4561)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient suffered from a fall and broke the distal interlocking screw after having a fracture fixation surgery of the right femur using the trigen tan system.Overall, the alignment remained intact, and the screw had no migration.The patient achieved union at 8 weeks after the primary surgery.No hardware revision surgery was scheduled, and the patient was lost to follow up.This incident was noticed in a retrospective post-market clinical follow up activity (pmcf) where anonymized data summarizing outcomes were gathered; therefore, additional information is not known, and it is not possible to collect it.We do not have the opportunity to identify individual hcps and revisit negative feedback to retrieve further information.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal reference number: (b)(4).This complaint was opened by smith+nephew to document a patient complication identified through a review of clinical evidence from a post market clinical follow up activity sponsored by smith and nephew.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).Part number updated to unknown trigen tan nail screw.Nail added as concomitant device.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient suffered from a fall on (b)(6) 2017 and broke the distal interlocking screw after having a fracture fixation surgery of the right femur using the trigen tan system on (b)(6) 2016.Overall, the alignment remained intact, and the screw had no migration.The patient achieved union at 8 weeks after the primary surgery.No hardware revision surgery was scheduled, and the patient was lost to follow up.This incident was noticed in a retrospective post-market clinical follow up activity (pmcf) where anonymized data summarizing outcomes were gathered; therefore, additional information is not known, and it is not possible to collect it.We do not have the opportunity to identify individual hcps and revisit negative feedback to retrieve further information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: a4, a6, b5, d1, d2a, d2b, d4, d10, h6 (health effect - clinical code).
|
| |
|
Additional Manufacturer Narrative
|
|
H3,h6: given the nature of the alleged incident, the devices could not be returned for evaluation; therefore, a device analysis could not be performed.However, the clinical/medical investigation concluded that the provided images confirm the reported broken screw.Based on a review of the information provided, the clinical root cause of the reported distal interlocking screw breakage was the reported fall sustained by the patient, the reported pain was a result of the broken screw.It cannot be concluded there was a mal performance of the implant or implant failure of the smith and nephew products.Devices' batch numbers were not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 31 months revealed similar events for the low profile screw and hex screw, and one similar event for the additional hex screw, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of complaint history for the previous 31 months did not reveal similar events for the nail.A review of the instructions for use documents for intramedullary nail system revealed that cracking or fracture of the implant may occur.Preoperative planning section states that proper type and size of implant must be selected to ensure effective treatment of patients considering factors such as patient¿s size, strength, skeletal characteristics, skeletal health, and general health.Overweight or musculoskeletal deficient or unhealthy patients may create greater loads on implants that may lead to breakage or other failure of the implants.Additionally, the postoperative care section warns that early weight bearing substantially increases implant loading and increases the risk of breaking the device.This has been identified as a possible adverse effect.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.Per material specification, the quality and manufacture of standard grade titanium, aluminum and vanadium alloy shall be controlled.At this time, we have no evidence to conclude that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include patient anatomy, abnormal loading of limb, excessive forces and/or traumatic injury.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|