• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS SODIUM ELECTRODE; ELECTRODE, ION SPECIFIC, SODIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS SODIUM ELECTRODE; ELECTRODE, ION SPECIFIC, SODIUM Back to Search Results
Catalog Number 21029371001
Medical Device Problem Codes Mechanical Problem (1384); Low Test Results (2458); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable ise indirect gen 2 sodium results for qc and patient samples from cobas integra 400 plus analyzer serial number 421575.The initial result was 129.9 mmol/l and the repeat result was 136.2 mmol/l.The questionable result was reported outside of the laboratory and was corrected with the repeat result.
 
Additional Manufacturer Narrative
The customer replaced the sodium electrode, ise indirect and direct calibrators, and ise solutions.The ise calibration and qc were acceptable.The field service engineer replaced the ise tubing and all three electrodes.The investigation is ongoing.
 
Additional Manufacturer Narrative
The investigation did not identify a product problem.The cause of the event could not be determined.The customer replaced all electrodes, ran precision testing, and ran comparisons on patient samples.Afterward, the customer had no further issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SODIUM ELECTRODE
Common Device Name
ELECTRODE, ION SPECIFIC, SODIUM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16014176
Report Number1823260-2022-04100
Device Sequence Number6561101
Product Code JGS
UDI-Device Identifier04015630032457
UDI-Public04015630032457
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K963627
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number21029371001
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer03/08/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date03/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-