| Model Number |
X SERIES |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device failed self-test for defib.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device was unable to detect the attached multifunction cable.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also correcting information submitted on the initial medwatch report.Please reference section b5 and h6 (medical device problem code).Device evaluation: zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing including power cycling and 30j self-test, using the returned multifunction cable, without duplicating the report.An internal inspection of the device found no discrepancies.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|