| Model Number |
ROB10027 |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during the set-up for a cori assisted surgery, a real intelligence tablet displayed a dvi input error which results that the cori unit turns on and off.Surgery was performed manually without any delay, since this was noticed during the set-up for surgery, the patient was not yet involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|