| Model Number |
7887-010R |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During patient positioning the levo dual motion base was slipping in the longitudinal direction while locked.It was especially noticeable when the patient was in reverse trendelenburg position.
|
| |
|
Additional Manufacturer Narrative
|
|
Following the evaluation of the returned device, it was found that the internal components (cams) were improperly installed during the assembly process and did not fully engage with the gear racks when the brake handles were in locked position.An additional space was created between the tooth locks and gear rack as a result of improper installation causing the longitudinal slip motion of the device.A replacement device with correct assembly/installation of the components was provided to the customer.
|
| |
|
Event or Problem Description
|
|
During patient positioning the levo dual motion base was slipping in the longitudinal direction while locked.It was especially noticeable when the patient was in reverse trendelenburg position.
|
| |
|