| Model Number |
H21911 |
| Medical Device Problem Code |
No Tactile Prompts/Feedback (4024)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, it was noted that the ep-4 stimulator was showing as off and not responding to touchscreen inputs which caused the procedure to be cancelled.There were no adverse consequences to the patient.
|
| |
|