• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. ALTIVATE ANATOMIC SHOULDER; ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 50, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. ALTIVATE ANATOMIC SHOULDER; ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 50, E-PLUS Back to Search Results
Model Number 521-07-250
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as conversion of reverse shoulder.The previous surgery and the surgery detailed in this event occurred 1 year and 5 months apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements and showed no non-conforming material reports (ncmr) associated with the main contributor component listed in the complaint.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to conversion of reverse shoulder.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient had failed djo anatomic.Surgeon explanted all implants and put in a djo reverse shoulder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIVATE ANATOMIC SHOULDER
Common Device Name
ALTIVATE ANATOMIC, ALL-POLY PEGGED GLENOID, SIZE 50, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key16015381
Report Number1644408-2022-01724
Device Sequence Number6419086
Product Code KWS
UDI-Device Identifier00190446178385
UDI-Public(01)00190446178385
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K162024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/03/2026
Device Model Number521-07-250
Device Catalogue Number521-07-250
Device Lot Number893U1176
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2022
Initial Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-07-000 LOT 888U1571; 520-10-000 LOT 851U1310; 520-54-318 LOT 959U1012
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age50 YR
Patient SexMale
-
-