• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. META RETRO FEM NAIL 11.5 X 42; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. META RETRO FEM NAIL 11.5 X 42; NAIL, FIXATION, BONE Back to Search Results
Model Number 71653242
Medical Device Problem Codes Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870); Failure of Implant (1924); Pain (1994)
Date of Event 08/20/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal reference number: (b)(4).This complaint was opened by smith+nephew to document a patient complication identified through a review of clinical evidence from a post market clinical follow up activity sponsored by smith and nephew.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that the patient had a breakage of the distal interlocking screws after having a fracture fixation surgery of the left femur using the trigen retrograde femoral nail system.Also, there was what appears to be a new femoral shaft fracture surrounding nail proximal to original fracture.The patient achieved union at 27 weeks after the primary surgery.No hardware revision surgery was scheduled.The current state of health of the patient is unknown.This incident was noticed in a retrospective post-market clinical follow up activity (pmcf) where anonymized data summarizing outcomes were gathered; therefore, additional information is not known, and it is not possible to collect it.We do not have the opportunity to identify individual hcps and revisit negative feedback to retrieve further information.
 
Additional Manufacturer Narrative
H10 (sharing the results of investigation): given the nature of the alleged incident, the device, could not be returned for evaluation.The clinical/medical investigation concluded that, the information provided was obtained from a post-market clinical follow up activity and was reviewed.However, as of the date of this medical investigation, no relevant supporting clinical information has been provided for evaluation.It is noted that further information will not be provided.Based on the information provided, the patient had a breakage of the distal interlocking screws post fracture fixation surgery of the left femur, in addition to, an apparent a new femoral shaft fracture surrounding nail proximal to original fracture.According to the report, the patient achieved union at 27 weeks after the primary surgery, and no hardware revision has been scheduled.Since it was reported the current health status is unknown, the impact to the patient beyond that which has already reported cannot be confirmed nor concluded.Therefore, no further clinical/medical assessment is warranted at this time.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.For the trigen intertan intertroch antegr nail screw, trigen tan/fan nail screw, trigen meta-tan troch antegr nail screw, specific identifiers were not provided, thus, a complaint history review could not be performed.A review of the instructions for use documents for intramedullary nail system revealed that care prior to bony union stablish to immobilize and/or externally support skeletal structures that have been implanted with surgical metallic implants until skeletal union is observed.Patients who are obese and/or noncompliant, as well as patients who could be pre-disposed to delayed or non-union, should have auxiliary support.Patients and nursing care providers should be advertised of these risks.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.With the results of this investigation the root cause of this event could not be determined.Factors that could contribute to the reported event include patient bone quality, post-operative patient condition and/or surgical technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H6 (corrected medical device problem code).
 
Event or Problem Description
It was reported that the patient had what appears to be a new femoral shaft fracture surrounding nail proximal to original fracture after having a fracture fixation surgery of the left femur using the trigen retrograde femoral nail system on (b)(6) 2020.The patient achieved union at 27 weeks after the primary surgery.No hardware revision surgery was scheduled.The current state of health of the patient is back to working part time and doing well.This incident was noticed in a retrospective post-market clinical follow up activity (pmcf) where anonymized data summarizing outcomes were gathered; therefore, additional information is not known, and it is not possible to collect it.We do not have the opportunity to identify individual hcps and revisit negative feedback to retrieve further information.
 
Additional Manufacturer Narrative
Additional information: a4, b3, b5, d10, h6 (health effect - clinical code).
 
Additional Manufacturer Narrative
H3, h6: given the nature of the alleged incident, the devices could not be returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, based on the pre/post implantation imaging, it cannot be ruled out whether the fracture was a result from failure to achieve adequate initial fixation which contributed to the new femoral shaft fracture surrounding nail proximal to original fracture.Although, at two and one-half months post-implantation full consolidation is not expected.Factors that could contribute to the reported event include noncompliant and the patient¿s continued use of alcoholic drinks, benzodiazepines and cocaine.The impact to the patient was the breakage of the two distal interlocking screws and a new femoral shaft fracture surrounding nail proximal to original fracture.Of note, the patient achieved union at 27 weeks after the primary surgery.Since no hardware revision surgery was scheduled and the patient is back to working part time and doing well, no further clinical/medical assessment is warranted at this time.The nail's batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of the low profile screws' production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the previous 31 months did not reveal similar events for the nail.A review of complaint history based on the historical data of the low profile screw revealed a similar event for the listed batch.A review of complaint history revealed a similar event for the additional low profile crew over the previous 31 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for intramedullary nail system revealed that care prior to bony union establish to immobilize and/or externally support skeletal structures that have been implanted with surgical metallic implants until skeletal union is observed.Patients who are obese and/or noncompliant, as well as patients who could be pre-disposed to delayed or non-union, should have auxiliary support.Patients and nursing care providers should be advertised of these risks.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the products failed to meet any specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
META RETRO FEM NAIL 11.5 X 42
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16015553
Report Number1020279-2022-04995
Device Sequence Number7303150
Product Code JDS
UDI-Device Identifier03596010548467
UDI-Public03596010548467
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K051557
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Other,Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71653242
Device Catalogue Number71653242
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2022
Supplement Date Received by Manufacturer01/13/2023
08/22/2024
10/14/2024
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date01/16/2023
08/23/2024
10/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71645035/TRIGEN LOW PROFILE SCREW 5.0MM X 35MM; 71645035/TRIGEN LOW PROFILE SCREW 5.0MM X 35MM
Outcome Attributed to Adverse Event Other;
Patient Age35 YR
Patient SexMale
Patient Weight109 KG
-
-