| Model Number |
E411 RACK |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The calibration and qc data were acceptable.Therefore general reagent issues were excluded.Liquid level detection alarms on the day the events indicate possible improper sample pre-analytics.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable results from the cobas e411 rack analyzer.Patient 1 initial total psa elecsys result was 0.02 ng/ml and was reported outside of the laboratory.The treating physician questioned the result as it did not match the patient's clinical history.On (b)(6) 2022, the repeat result was 1.29 ng/ml.On (b)(6) 2022, patient 2 initial insulin elecsys result was 0.04 uu/ml and was not reported outside of the laboratory.The repeat result was 27 uu/ml.The total psa reagent lot number was 648973 with an expiration date of 31-jan-2024.The insulin reagent lot number was 586607 with an expiration date of 31-mar-2023.
|
| |
|