ROCHE DIAGNOSTICS ONLINE TDM VANCOMYCIN GEN.3; VANCOMYCIN IMMUNOASSAY Back to Search Results
Catalog Number 08445605190
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable vanc2 vancomycin results for 2 patient samples on a cobas pro c 503 analytical unit.The initial results were reported outside of the laboratory.The customer was prompted to repeat the samples after qc was performed and was outside of the acceptable range.For patient 1, the initial vanc2 result was 11.9 ug/ml.The sample was repeated in duplicate on another analyzer and the results were 9.2 ug/ml and 9.5 ug/ml.For patient 2, the initial vanc2 result was 10.6 ug/ml.The sample was repeated in duplicate on another analyzer and the results were 8.3 ug/ml and 8.0 ug/ml.The repeat results were deemed correct.The analyzer serial number is (b)(4).
 
Additional Manufacturer Narrative
Qc was within range prior to the event.The investigation is ongoing.
 
Additional Manufacturer Narrative
Calibration was last performed on (b)(6) 2023.The field service engineer (fse) replaced syringe seals and the heat cut filter, flushed the sample and reagent probes, and checked the rinse levels.The investigation did not identify a product problem.The root cause of the event could not be determined.The service maintenance actions performed by the fse resolved the issue.No further issues were reported after the service visit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONLINE TDM VANCOMYCIN GEN.3
Common Device Name
VANCOMYCIN IMMUNOASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16016297
Report Number1823260-2022-04110
Device Sequence Number6738350
Product Code LEH
UDI-Device Identifier07613336157800
UDI-Public07613336157800
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152245
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Catalogue Number08445605190
Device Lot Number64125401
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/30/2022
Supplement Date Received by Manufacturer02/07/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date03/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-