Sections d9, h3 and imdrf codes were updated.1 sample was provided for investigation.An interfering factor against streptavidin agent could be identified in this sample, as changing signal values were observed for the a-tg v1 vs.A-tg v2 reagent.Customer¿s values for ft3 iii and ft4 3 were reproduced.An influence of prewash could be observed for the ft3 iii v1 assays.The rarely occurring event of the detected interfering factor is covered by a disclaimer in the section limitation - interference in the method sheets of all applicable products: in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.A product problem could not be found.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.
|