MAUDE Adverse Event Report: BTG YOKNEAM ICEROD PLUS 90 DEGREE NEEDLE; UNIT, CRYOSURGICAL, ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BTG YOKNEAM ICEROD PLUS 90 DEGREE NEEDLE; UNIT, CRYOSURGICAL, ACCESSORIES Back to Search Results
Model Number FPRPR3508
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that frost was observed on the needle shaft.This icerod plus 90 degree needle was selected for a percutaneous cryoablation procedure for residual tumor after tace for liver cancer.During the testing phase, the whole needle was found to be covered with frost.The needle was not used for the procedure; no patient complications were reported.
 
Additional Manufacturer Narrative
Device eval by manufacturer: this icerod plus 90 degree needle was returned for analysis.Visual inspection of the exterior of the needle revealed no abnormalities.The needle was connected to the icefx console, and a two-minute confidence test was performed without any abnormal functionality.Further testing included a five-minute freeze cycle, and it was found that the needle shaft became completely covered in frost.The needle was disassembled, and the vacuum sleeve was tested for vacuum integrity.The vacuum sleeve became covered in frost, indicating the vacuum sleeve was defective.The vacuum sleeve was inspected under a microscope for abnormalities revealing discoloration and abnormalities that likely led to the vacuum integrity failing.
 
Event or Problem Description
It was reported there was frost observed on the needle shaft.This icerod plus 90 degree needle was selected for a percutaneous cryoablation procedure for residual tumor after tace for liver cancer.During the testing phase, the whole needle was found to be covered with frost.The needle was not used for the procedure; no patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICEROD PLUS 90 DEGREE NEEDLE
Common Device Name
UNIT, CRYOSURGICAL, ACCESSORIES
Manufacturer (Section D)
BTG YOKNEAM
tavor building 1
po box 224
yokneam 20692 03
IS  2069203
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
tavor building number 1, yokne
yokneam 20692 03
IS   2069203
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16016896
Report Number2124215-2022-52902
Device Sequence Number6570449
Product Code GEH
UDI-Device Identifier07290104830563
UDI-Public07290104830563
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110946
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/17/2024
Device Model NumberFPRPR3508
Device Catalogue NumberFPRPR3508
Device Lot Number0029734560
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/05/2022
Supplement Date Received by Manufacturer01/11/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
Patient Weight70 KG
-
-