MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-15-T; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-15-T; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 133S-MX-15-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Seroma (2069)
Date of Event 11/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "left breast tissue expander was quite a bit softer than it had been", "the volume seemed to have decreased", "ruptured left breast tissue expander", and "some serous fluid within the tissue expander and it was noted to have failed at a posterior seam of the 2 piece shell, at about the 1 o'clock position when viewed anteriorly".Device has been explanted and replaced.
 
Additional Manufacturer Narrative
(b)(4).In response to fda report number: (b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.The event of seroma is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation and seroma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATRELLE TE SMOOTH 133S-MX-15-T
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key16017075
Report Number9617229-2022-22653
Device Sequence Number17784521
Product Code LCJ
UDI-Device Identifier10888628044210
UDI-Public10888628044210
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K182054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/07/2025
Device Model Number133S-MX-15-T
Device Catalogue Number133S-MX-15-T
Device Lot Number3419825
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2022
Initial Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexFemale
Patient Weight92 KG
Patient EthnicityHispanic
-
-