Catalog Number 56535RIM
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The account alleges that during an interventional percutaneous transhepatic biliary drainage/stenting procedure, the 4f catheter tip detached within the patient when the physician was trying to maneuver the catheter across the biliary artery.The foreign body was left behind within the patient's vascular system.The patient is scheduled for an additional foreign body removal procedure.
 
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow-up report will be submitted once the evaluation has been complete.
 
Additional Manufacturer Narrative
The suspect medical device was not returned for investigation.The complaint could not be confirmed.The root cause could not be determined.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Should the device be returned later, the investigation will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED CATHETER
Common Device Name
PERIPHERAL CATHETERS
Manufacturer (Section D)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer (Section G)
MERIT MEDICAL MANUFACTURING
14646 kirby dr
houston TX 77047
Manufacturer Contact
bryson heaton bsn,rn.
1600 merit parkway
south jordan, UT 84095
MDR Report Key16018561
Report Number3010665433-2022-00128
Device Sequence Number5896401
Product Code DQO
UDI-Device Identifier00884450020135
UDI-Public884450020135
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K053171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number56535RIM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2022
Supplement Date Received by Manufacturer03/10/2023
Initial Report FDA Received Date12/20/2022
Supplement Report FDA Received Date03/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDWIRE
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-